Experimental Studies on the Capillary Flow Phenomena in Polyimide Based Glass Microfluidic Devices

Authors

  • Subhadeep Mukhopadhyay

Keywords:

SU-8, Polyimide, Capillary flow, Micropillar, Surface to volume ratio

Abstract

The polyimide based glass microfluidic devices are fabricated using the maskless lithography. Water is the only working liquid in this work. The effects of micropillar side length, microfluidic friction and surface area to volume ratio on the surface-driven capillary flow of water are studied experimentally. The surface-driven capillary flow of water is slower due to higher microfluidic friction. This work will be useful to control the water inside the microfluidic lab-on-a-chip systems for commercial applications.

Cite this Article
Subhadeep Mukhopadhyay. Experimental Studies on the Capillary Flow Phenomena in Polyimide Based Glass Microfluidic Devices. Emerging Trends in Chemical Engineering. 2017; 4(3): 22–26p.

Author Biography

  • Subhadeep Mukhopadhyay

    Department of Electronics and Computer Engineering, National Institute of Technology Arunachal
    Pradesh, Yupia, Papum Pare, Arunachal Pradesh, India

Published

2018-01-08

Issue

Section

Articles